Adjust To Invial Pusher Position - Thermo Scientific VF-F20-A Operating Manual

Integral fraction collector ft
Table of Contents

Advertisement

6.8.2

Adjust to InVial Pusher Position

Preparations
Follow these steps
Integral Fraction Collector FT (VF-F20-A) Operating Manual
1. Open the left front
2. Press the service button on keypad to move the needle unit into
service position.
NOTICE To avoid personal injury, do not put hands into the
compartment when the needle unit is moving.
1. Raise the lock pin upwards by hand. The Lock tab will pop out from
the U-shape slot accordingly.
TIP Do not press the lock pin. Just lift it up.
Figure 55: Moving the lock pin up by hand
2. Release the lock pin. The pusher is released to InVial position.
3. Physically check that the needle tip does not protrude out of the
pusher and close the left front door.
door.
6
Operation
Page 135