Thermo Scientific VF-F20-A Operating Manual page 211

Integral fraction collector ft
Table of Contents

Advertisement

8
Troubleshooting
Message and Code
Description and Remedial Action
Code 9060.
Invalid position at the end of a collection (injection with Fraction
Collection not off) with Tube Wrapping disabled in the instrument
method. This warning indicates that the current collection has finished
successfully, but the last available tube has been used. Enable Tube
Wrapping or replace the racks and reset the Tube Position on ePanel
FC tab, otherwise, users will not be able to start another collection. An
error "Maximum number of installed tubes used up" will be popped up
during the injection ready check.
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 211