Maintenance and Service
Page 198
2. Push all tubing and capillaries, which are present in the tubing chase
of the Vanquish system modules, into the tubing chase. Otherwise,
you will not be able to remove the slide-in module properly from the
enclosure in the next step.
3. Grasp the slide-in module by the middle frame where the door lock
is placed and pull the module out of the enclosure by approximately
10 cm.
Figure 86: Pulling out the slide-in module
NOTICE
The slide-in module can fall down when pulling it out of the
enclosure too far. Pull out the slide-in module just far enough so
that you can grasp it on both sides from below.
4. Remove the slide-in module from the enclosure.
The following steps require a team effort:
a) Take the slide-in module on both sides from below.
b) Pull the slide-in module from the rails towards the front.
c) Place the slide-in module on a clean and stable surface.
5. Return the slide-in module:
a) To request a dedicated packaging for the slide-in module and for
the appropriate procedure for returning the module, contact
your local Thermo Fisher Scientific support organization.
b) Pack the slide-in module in the dedicated packaging.
Integral Fraction Collector FT (VF-F20-A) Operating Manual