Integral Fraction Collector FT (VF-F20-A) Operating Manual
Figure 25: Flow path connection in a two stacks configuration
For a two stacks configuration with a Vanquish Horizon or Flex UHPLC
system the flow path connection are as follows:
No.
Description
1
0.10 x 350 mm, MP35N
2
Active or passive pre-heater, MP35N
3
Depending on flow rate:
Up to 2 mL/min: 0.1 x 550 mm, MP35N
2 - 10 mL/min: 0.18 x 550 mm, MP35N
4
Depending on fraction collection mode
and flow rate
Delay capillary for time-based
fractionation:
Up to 2 mL/min: 0.1 x 350 mm, MP35N
2 - 10 mL/min: 0.18 x 350 mm, MP35N
Delay capillaries for peak-based
fractionation
Up to 0.5 mL/min: 0.18 x 1200 mm,
MP35N
0.5 – 1 mL/min: 0.25 x 1500 mm, MP35N
1 - 2 mL/min: 0.5 x 800 mm, PEEK
2 - 10 mL/min: 1 x 1000 mm, PEEK
5
Installation
Shipped with
System Base Vanquish Horizon/Flex
System Base Vanquish Horizon/Flex
Integral Fraction Collector FT
Integral Fraction Collector FT
Optional, needs be ordered separately
Page 75