5
Installation

5.11 Setting Up the Device in the Software

Check the setting of the needle capillary
Follow these steps
Page 94
This manual assumes that the chromatography software is already
installed on the data system computer and a valid license is available.
For more information about setting up the Vanquish system in the
software, refer to the Vanquish System Operating Manual.
The Help for the software that you are using provides detailed
information about the settings on each property page.
The NeedleCapillary property must be set correctly in the software
according to needle capillary installed on device, see
Capillaries, and Tubing
default shipment is ID 0.18 mm. This is also the default setting for
needle capillary in the Chromeleon.
If the needle capillary with ID 0.25 mm is installed the new capillary
dimensions must be set in Chromeleon accordingly.
1. Check the marker physically located on the needle capillary. It is
either ID 0.18 mm or ID 0.25 mm. The default needle capillary
installed in the device is ID 0.18 mm.
2. In Chromeleon Console click Command icon or press F8.
Figure 38: Selecting the Command configuration page
Integral Fraction Collector FT (VF-F20-A) Operating Manual
(
page
71). The needle capillary included in the
Connecting Fittings,