Maintenance and Service
7.5.2

Securing the Needle Drive

When
Page 154
6. Press the Service button on the keypad to move the needle unit to
service position and then press the service button again to move the
needle unit to parking position.
Before maintenance procedures or for transport, make sure to secure
the needle drive.
NOTICE
If the needle drive is not secured during maintenance or transport, it can
damage the device. Secure the needle drive when instructed to do so to
avoid damage to the device.
When the device is powered off, the needle drive may move inside the
fraction collector compartment when the device is tilted, moved from
one place to another place.