Connecting The Delay Capillary - Thermo Scientific VF-F20-A Operating Manual

Integral fraction collector ft
Table of Contents

Advertisement

5
Installation
Purging the Flush Pump
5.9.3

Connecting the Delay Capillary

Parts required
Follow these steps
Page 90
6. Check the flush solvent tubing across the entire flow path:
a) Make sure the tubing is not bent, pinched or squeezed at any
point in the flow path.
b) If you have to cut tubing to length, use a tubing cutter. Make
sure that the cut is at right angle to the length of the line.
After the device has been turned on, purge the flush pump using
Chromeleon to fill the flush pump with the fresh flush solvent.
During purging, the flush pump and flush buffer loop are washed
continuously until all flush purge lines are filled with the fresh flush
solvent.
Delay Capillary
1. Remove the detector waste line if it has been installed onto detector
flow cell outlet or the outlet (OUT) on the flow cell connection unit.
Figure 36: Removing the detector waste line
No.
Description
1
Detector waste line
Integral Fraction Collector FT (VF-F20-A) Operating Manual