Thermo Scientific VF-F20-A Operating Manual page 83

Integral fraction collector ft
Table of Contents

Advertisement

Connecting the Needle Wash Inlet Line to the Needle Wash Reservoir
Integral Fraction Collector FT (VF-F20-A) Operating Manual
No.
Description
A
Guide hole
1+2
Not to be used for wash liquid line; reserved for other tubing
3
Tubing guide for wash liquid line
4
Not to be used for wash liquid line; reserved for other tubing
1. Connect the needle wash reservoir to the line:
a) Rinse the needle wash reservoir thoroughly with a high-purity
solvent.
b) Fill the needle wash reservoir with fresh needle wash liquid.
c) Feed the needle wash line through the retaining guide and
through an open hole in the cap of the needle wash reservoir.
The retaining guide keeps the tubing in place in the reservoir.
Close any open holes in the reservoir cap with cap plugs.
Figure 30: Securing the needle wash line with the reservoir cap
No.
Description
1
Needle wash inlet line
2
Retaining guide (1/8'')
3
Reservoir cap
4
Cap plugs
2. Fill the needle wash reservoir with needle wash liquid. Mind the
requirements outlined in the previous section.
5
Installation