Thermo Scientific VF-F20-A Operating Manual page 176

Integral fraction collector ft
Table of Contents

Advertisement

7
Maintenance and Service
Page 176
Figure 77: Route and secure the needle capillary on the front surface of
the device
NOTICE Do NOT over-tighten the attachment screw because the needle
capillary under the press plate may be damaged.
10. Install back the front plate with reverse steps. See
Front Plate
(
page 155).
11. Power on the device. Close the compartment doors and the device
will run initialization automatically. Make sure no errors or warnings
are reported in the Audit trail.
12. In Chromeleon, configure the Needle Capillary. There are two
options of the needle capillary types: 0.18 x 415 mm and 0.25 x 415
mm. Users can check via the marker on the needle capillary and
select accordingly. For more details, see
the Software
(
page
13. In Chromeleon, select Internal Needle Rinse on ePanel to rinse the
internal surface of the needle capillary and needle unit with mobile
phase. For more details, see
(
page 113).
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Setting Up the Device in
94).