Manual Pre-Cleaning With Ultrasound And Brush - Braun Aesculap TSPACE PEEK Instructions For Use/Technical Description

Hide thumbs Also See for Aesculap TSPACE PEEK:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
en

3.10.2 Manual pre-cleaning with ultrasound and brush

Phase
Step
I
Ultrasonic cleaning
II
Rinsing
D–W:
Drinking water
RT:
Room temperature
*Recommended: BBraun Stabimed fresh
Note the information on appropriate cleaning
brushes and disposable syringes, see Validated
cleaning and disinfection procedure.
Phase I
Clean the product in an ultrasonic cleaning bath
(frequency 35 kHz) for at least 15 min. Ensure that
all accessible surfaces are immersed and acoustic
shadows are avoided.
Clean the product with a suitable cleaning brush in
the solution until all discernible residues have been
removed from the surface.
If applicable, brush through non-visible surfaces
with an appropriate cleaning brush for at least
1 min.
Mobilize non-rigid components, such as set screws,
links, etc. during cleaning.
Thoroughly rinse through these components with
the cleaning disinfectant solution (at least five
times), using a disposable syringe.
16
T
t
Conc.
[°C/°F]
[min]
[%]
RT (cold)
>15
2
RT (cold)
1
-
Phase II
Water
Chemical
quality
D–W
Aldehyde-free, phenol-free, and
QUAT-free concentrate, pH ~ 9*
D–W
-
Rinse/flush the product thoroughly (all accessible
surfaces) under running water.
Mobilize non-rigid components, such as set screws,
joints, etc. during rinsing.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aesculap tspace xpAesculap tspace 3d

Table of Contents