3.2

General information

Dried or affixed surgical residues can make cleaning
more difficult or ineffective and lead to corrosion.
Therefore the time interval between application and
processing should not exceed 6 h; also, neither fixating
pre-cleaning temperatures >45 °C nor fixating disin-
fecting agents (active ingredient: aldehydes/alcohols)
should be used.
Excessive measures of neutralizing agents or basic
cleaners may result in a chemical attack and/or to fad-
ing and the laser marking becoming unreadable visu-
ally or by machine for stainless steel.
Residues containing chlorine or chlorides e.g. in surgi-
cal residues, medicines, saline solutions and in the ser-
vice water used for cleaning, disinfection and steriliza-
tion will cause corrosion damage (pitting, stress
corrosion) and result in the destruction of stainless
steel products. These must be removed by rinsing thor-
oughly with demineralized water and then drying.
Additional drying, if necessary.
Only process chemicals that have been tested and
approved (e.g. VAH or FDA approval or CE mark) and
which are compatible with the product's materials
according to the chemical manufacturers' recommen-
dations may be used for processing the product. All the
chemical manufacturer's application specifications
must be strictly observed. Failure to do so can result in
the following problems:
Optical changes of materials, e.g. fading or discol-
oration of titanium or aluminum. For aluminum,
the application/process solution only needs to be of
pH >8 to cause visible surface changes.
Material damage such as corrosion, cracks, fractur-
ing, premature aging or swelling.
Do not use metal cleaning brushes or other abra-
sives that would damage the product surfaces and
could cause corrosion.
Further detailed advice on hygienically safe and
material-/value-preserving reprocessing can be
found at www.a-k-i.org, link to "AKI-Brochures",
"Red brochure".
3.3

Reusable products

There is no set maximum number of uses and prep-
aration cycles for the product.
The life of the product is limited by damage, normal
wear and tear, type and duration of use, as well as
handling, storage and transport of the product.
A careful visual and functional inspection before
the next use is the best opportunity to recognize a
product that is no longer functional.
3.4

Preparations at the place of use

If applicable, rinse non-visible surfaces preferably
with deionized water, with a disposable syringe for
example.
Remove any visible surgical residues to the extent
possible with a damp, lint-free cloth.
Transport the dry product in a sealed waste con-
tainer for cleaning and disinfection within 6 hours.
3.5

Preparing for cleaning

Disassemble the product prior to cleaning, see Dis-
assembly.
For mechanical cleaning: mount the irrigation
connector 4 to the rear end of the rotary nut 3.
3.6

Disassembly

SN305R insertion instrument
Turn the rotary nut 3 counterclockwise ("loosen"
direction).
Turn the switch 2 to the left (direction
Fig. 1.
Turn the rotary nut 3 clockwise ("tighten" direc-
tion), until the rod 6 is out of the thread.
Remove the rod 6 from the tube 1, see Fig. 12.
1
6
Fig. 12 Insertion instrument disassembled
en
), see
9