2.3.3 Inserting the implant in the implant
bed
Picking up the implant with the insertion instru-
ment
Select an implant according to the size of the trial
implant.
Remove the implant with protective packaging
from the sterile packaging.
Mount the implant a in the protective packaging
onto the dihedron 7 of the insertion instrument A,
see Fig. 7. When doing so, pay attention to the ori-
entation of the implant (orientation according to
the marking on the insertion instrument).
a
a
7
Fig. 7
Coupling the implant to the insertion instru-
ment
Legend
a Implant
Turn the switch 2 to the right (direction
Fig. 3.
Hold the implant at 0° position, see Fig. 8.
a
Fig. 8
Implant at 0° position
Turn the rotary nut 3 on the insertion instrument A
clockwise (:"tighten" direction) until the trial
implant is clamped, see Fig. 4.
Remove the protective packaging from the implant.
Check manually that the implant is securely in
place.
A
), see
7
A
Filling the implant with bone material (optional)
Place implant a in filling holder E, see Fig. 9. When
doing so, observe the implant length (marking on
the filling holder: L26/L30/L34).
Fig. 9
Implant in filling holder
Compress the bone grafts or bone replacement
material in the inner window of the implant a using
the pestle provided in the set. When doing so,
ensure that the pestle does not damage the surface
of the implant.
Inserting the implant
Insert
the
implant a
instrument A straight (0° position) into the pre-
pared implant bed. To do so, carefully hit with a
hammer on the impact surface of the rotary nut 3.
A
Fig. 10 Inserting the implant into the implant bed
E
a
with
the
insertion
a
en
a
A
7