Manual Cleaning With Immersion; Disinfection - Braun Aesculap TSPACE PEEK Instructions For Use/Technical Description

Hide thumbs Also See for Aesculap TSPACE PEEK:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
en
3.8.1 Manual cleaning with immersion disinfection
Phase
Step
I
Disinfecting
cleaning
II
Intermediate rinse
III

Disinfection

IV
Final rinse
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)
RT:
Room temperature
*Recommended: BBraun Stabimed fresh
Note the information on appropriate cleaning
brushes and disposable syringes, see Validated
cleaning and disinfection procedure.
Phase I
Fully immerse the product in the cleaning/disinfec-
tant for at least 15 min. Ensure that all accessible
surfaces are moistened.
Clean the product with a suitable cleaning brush in
the solution until all discernible residues have been
removed from the surface.
If applicable, brush through non-visible surfaces
with an appropriate cleaning brush for at least
1 min.
Mobilize non-rigid components, such as set screws,
links, etc. during cleaning.
Thoroughly rinse through these components with
the cleaning disinfectant solution (at least five
times), using a disposable syringe.
Phase II
Rinse/flush the product thoroughly (all accessible
surfaces) under running water.
Mobilize non-rigid components, such as set screws,
joints, etc. during rinsing.
Drain any remaining water fully.
12
T
t
Conc.
[°C/°F]
[min]
[%]
RT (cold)
>15
2
RT (cold)
1
-
RT (cold)
5
2
RT (cold)
1
-
RT
-
-
Water
Chemical
quality
D–W
Aldehyde-free, phenol-free, and
QUAT-free concentrate, pH ~ 9*
D–W
-
D–W
Aldehyde-free, phenol-free, and
QUAT-free concentrate, pH ~ 9*
FD-W
-
-
-
Phase III
Fully immerse the product in the disinfectant solu-
tion.
Mobilize non-rigid components, such as set screws,
joints, etc. during rinsing.
Rinse lumens at least 5 times at the beginning of
the exposure time using an appropriate disposable
syringe. Ensure that all accessible surfaces are
moistened.
Phase IV
Rinse/flush the product thoroughly (all accessible
surfaces).
Mobilize non-rigid components, such as set screws,
joints, etc. during final rinse.
Rinse lumens with an appropriate disposable
syringe at least five times.
Drain any remaining water fully.
Phase V
Dry the product in the drying phase with suitable
equipment (e.g. cloth, compressed air), see Vali-
dated cleaning and disinfection procedure.