3.8.2 Manual cleaning with ultrasound and immersion disinfection

Phase
Step
I
Ultrasonic cleaning
II
Intermediate rinse
III
Disinfection
IV
Final rinse
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)
RT:
Room temperature
*Recommended: BBraun Stabimed fresh
Note the information on appropriate cleaning
brushes and disposable syringes, see Validated
cleaning and disinfection procedure.
Phase I
Clean the product in an ultrasonic cleaning bath
(frequency 35 kHz) for at least 15 min. Ensure that
all accessible surfaces are immersed and acoustic
shadows are avoided.
Clean the product with a suitable cleaning brush in
the solution until all discernible residues have been
removed from the surface.
If applicable, brush through non-visible surfaces
with an appropriate cleaning brush for at least
1 min.
Mobilize non-rigid components, such as set screws,
links, etc. during cleaning.
Thoroughly rinse through these components with
the cleaning disinfectant solution (at least five
times), using a disposable syringe.
Phase II
Rinse/flush the product thoroughly (all accessible
surfaces) under running water.
Mobilize non-rigid components, such as set screws,
joints, etc. during rinsing.
Drain any remaining water fully.
D
t
Conc.
[°C/°F]
[min]
[%]
RT (cold)
>15
2
RT (cold)
1
-
RT (cold)
5
2
RT (cold)
1
-
RT
-
-
Water
Chemical
quality
D–W
Aldehyde-free, phenol-free, and
QUAT-free concentrate, pH ~ 9*
D–W
-
D–W
Aldehyde-free, phenol-free, and
QUAT-free concentrate, pH ~ 9*
FD-W
-
-
-
Phase III
Fully immerse the product in the disinfectant solu-
tion.
Mobilize non-rigid components, such as set screws,
joints, etc. during rinsing.
Rinse lumens at least five times at the beginning of
the exposure time with an appropriate disposable
syringe. Ensure that all accessible surfaces are
moistened.
Phase IV
Rinse/flush the product thoroughly (all accessible
surfaces) under running water.
Mobilize non-rigid components, such as set screws,
joints, etc. during final rinse.
Rinse lumens with an appropriate disposable
syringe at least five times.
Drain any remaining water fully.
Phase V
Dry the product in the drying phase with suitable
equipment (e.g. cloth, compressed air), see Vali-
dated cleaning and disinfection procedure.