Percent Oxygen Analyzer
If setting or Interval of days is changed, press the floppy disk icon
to save the setting. Changing the interval of days and saving it will
automatically reset the counter of days remaining.
4.10
No Sensor Detection Feature, Sensor Fail
This feature tries to detect whether a sensor is installed in the
analyzer or not. The software checks if the output of the amplifier is
inside a window around electronic zero level (output of the amplifier
without sensor). If it is, a 30 minute counter starts at the end of which a
warning is displayed and system alarm is triggered. If output rises above
the electronic zero detection window, during the countdown, the counter
is reset back to 30 minutes.
To set the feature the operator must find the "Sensor Fail" function
in the system menu.