DIAGNOSTICS
10DEC05
14006 - DUPLEX CAM: Initialization Error due to the PLATE PRESENT
DG4825-1
Page
SENSOR S5 reading blocked when the SLOW SCAN AY is in the home
88 of 180
position
Cause 1: The PLATE PRESENT SENSOR S5 operates correctly, but another malfunction
Action:
Cause 2: The PLATE PRESENT SENSOR S5 malfunctioned.
Action:
14008 - SLOW SCAN: Initialization Error
Cause:
Action:
14009 - DUPLEX CAM: Initialization Error
Cause:
Action:
14010 - Slow Scan Async Reset Error
Cause:
Action:
has occurred. This error occurs when initializing, either when the CR 825/850
SYSTEM is energized or when a mechanical subsystem initializes.
Check:
•
PLATE was removed from the CASSETTE
•
SLOW SCAN MOTOR operates correctly
•
obstruction in the path of the EXTRACTION BAR
Do the checkout
PLATE PRESENT SENSOR
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Do the checkout for the
BOARD
A7.
The DUPLEX CAM malfunctioned.
Do the checkout for the
The CR 825/850 SYSTEM sent a message to reset the SLOW SCAN AY.
None.
SLOW SCAN CONTROLLER BOARD A6 and COIL
CASSETTE HANDLING
S5.
AY.
Error Codes