SERVICE GUIDE
Replacing the Plate Size Sensor
Laser Warning
To avoid possible eye damage, before operating the unit without the cover, disconnect the laser (see
the Laser" on Page
35). For additional laser precautions see
personnel may remove the cover.
[1] Remove the main cover (see
Plate Size Sensor (1 of 4)
[2] Locate the Sensor and remove the 2 Allen screws.
[3] Cut the wire ties holding the sensor cable, and follow the cable up to the sensor board.
[4] Disconnect the cable from the sensor board.
[5] Reverse the above steps to install the plate size sensor, securing the cable with a new wire tie.
Plate Guide Passing Under Plate Size Sensor
[6] After installation, make sure that the plate guide passes under the sensor without interference.
62
"Removing the Main Cover" on Page
"Laser Safety Instructions" on Page
33).
10APR2006 – Catalog No. 6H1344
"Disconnecting
11. Only authorized