10008 - Cassette Rear Sensor S3 Did Not Clear - Kodak DirectView CR 825 Diagnostics Manual

Hide thumbs Also See for DirectView CR 825:
Table of Contents

Advertisement

DIAGNOSTICS
10DEC05
Cause 2: The CASSETTE REAR SENSOR S3 malfunctioned.
DG4825-1
Page
Action:
59 of 180

10008 - CASSETTE REAR SENSOR S3 did not clear

Cause 1: The CASSETTE REAR SENSOR S3 operates correctly, but another malfunction
Action:
Cause 2: The CASSETTE REAR SENSOR S3 malfunctioned.
Action:
Check:
status of CASSETTE REAR SENSOR S3 - use
Control
LED DS22 on the MOTION SYSTEM CONTROL BOARD A1 changes status
power to the CASSETTE REAR SENSOR S3
condition of the CASSETTE REAR SENSOR S3
occurred.
Check:
DRIVE ROLLERS rotate freely
CASSETTE REAR SENSOR S3 for obstructions
adjustment of the
Check:
status of the CASSETTE REAR SENSOR S3 - use
Control
LED DS22 on the MOTION SYSTEM CONTROL BOARD A1 changes status
power to the CASSETTE REAR SENSOR S3
condition of the CASSETTE REAR SENSOR S3
SLED CAM FOLLOWER