DIAGNOSTICS
10DEC05
DG4825-1
Page
135 of 180
SENSOR
S9 FLAG
H177_3023GCC
H177_3023GC
4
Measure the voltage at the EXTRACTION BAR SENSOR S9:
Negative
Green wire
5
If the voltage is not correct, check the condition of the HARNESS for the CASSETTE
HANDLING AY.
6
Do the checkout procedure for the
EXTRACTION BAR
SENSOR S9
EXTRACTION BAR
SLOW SCAN AY
MULTIMETER
Positive
White wire
Not Blocked
3
Check:
EXTRACTION BAR SENSOR
Status
Blocked
5 V DC
< 0.5 V DC
MOTION SYSTEM CONTROL BOARD
Checkout Procedures
SLOW SCAN AY moves:
down for 3 seconds
up to the home position
EXTRACTION BAR HOME
POSITION
condition of:
EXTRACTION BAR SENSOR S9
SENSOR S9 FLAG
S9
Illuminated
Not illuminated
LED DS28
A1.