16016 - Malfunctioning Pmt; 16017 - Possible Light Leak; 16019 - Calibration Could Not Find Offset, Rotate Galvo Counterclockwise; 16020 - Calibration Could Not Find Offset, Rotate Galvo Clockwise - Kodak DirectView CR 825 Diagnostics Manual

Hide thumbs Also See for DirectView CR 825:
Table of Contents

Advertisement

DIAGNOSTICS
10DEC05

16016 - Malfunctioning PMT

DG4825-1
Page
Description: A PMT does not match the requirements of the diagnostic test.
99 of 180
Cause:
Action:

16017 - Possible Light Leak

Cause:
Action:
16018 - Possible Light Leak
Description: An error occurred during "AUTO-ZERO." The PMTs detected a large
Cause:
Action:

16019 - Calibration could not find offset, rotate GALVO counterclockwise

Cause:
Action:

16020 - Calibration could not find offset, rotate GALVO clockwise

Cause:
Action:
The PMT/DAS BOARD A5 malfunctioned.
Do the checkout for the
The DOOR is open, a PANEL is loose, or a light leakage occurred in the
CASSETTE HANDLING SYSTEM.
1.
Do the
Light Test
2.
Repair the light leakage.
quantity of light.
The DOOR is open, a PANEL is loose, or a light leakage occurred in the
CASSETTE HANDLING SYSTEM.
1.
Do the
Light Test
2.
Repair the light leakage.
The calibration could not locate the "Offset."
Rotate the GALVO counterclockwise.
The calibration could not locate the "Offset."
Rotate the GALVO clockwise.
PMT/DAS BOARD
to locate the source of the light leakage.
to locate the source of the light leakage.
A5.
Error Codes