DIAGNOSTICS
10DEC05
12914 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
DG4825-1
Page
83 of 180
Description: The command MOVE TO EXTRACTION BAR REFERENCE SENSOR
Cause:
Action:
12915 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
Description: The command MOVE TO THE PLATE GUIDE POSITION did not occur.
Cause:
Action:
12930 - SLOW SCAN: Unexpected Retry during Current Test
Description: The system was not able to move to the position selected in the
Cause:
Action:
12931 - SLOW SCAN: Unexpected Retry during Velocity Test
Description: The system was not able to move to the position selected in the
Cause:
Action:
did not occur. The command is executed again.
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Do the checkout for the
BOARD
A7.
The command is executed again.
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Do the checkout for the
BOARD
A7.
diagnostics. The command is executed again.
The SLOW SCAN CONTROLLER BOARD malfunctioned.
Do the checkout for the
BOARD
A7.
diagnostics. The command is executed again
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Do the checkout for the
BOARD
A7.
SLOW SCAN CONTROLLER BOARD A6 and COIL
SLOW SCAN CONTROLLER BOARD A6 and COIL
SLOW SCAN CONTROLLER BOARD A6 and COIL
SLOW SCAN CONTROLLER BOARD A6 and COIL
Error Codes