DIAGNOSTICS
10DEC05
12941 - Unexpected Retry during Communication to MOTION SYSTEM
DG4825-1
Page
CONTROL BD
84 of 180
Description: A message could not be sent to the SLOW SCAN CONTROLLER
Cause:
Action:
ERASE LAMPS
13001 - 13008 ERASE LAMP PAIRS
Code
13001
13002
13003
13004
13005
13006
13007
13008
Cause:
Action:
BOARD A6. The command is executed again.
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Do the checkout for the
BOARD
A7.
Log Message
ERASE LAMP PAIR No. 1 is out
ERASE LAMP PAIR No. 2 is out
ERASE LAMP PAIR No. 3 is out
ERASE LAMP PAIR No. 4 is out
ERASE LAMP PAIR No. 5 is out
Two ERASE LAMP PAIRS are not functioning
Three ERASE LAMP PAIRS are not functioning
Four ERASE LAMP PAIRS are not functioning
One or more ERASE LAMP PAIRS malfunctioned.
Install a new ERASE LAMP PAIR.
SLOW SCAN CONTROLLER BOARD A6 and COIL
Error Codes