12941 - Unexpected Retry During Communication To Motion System; Control Bd; Erase Lamps; 13001 - 13008 Erase Lamp Pairs - Kodak DirectView CR 825 Diagnostics Manual

Hide thumbs Also See for DirectView CR 825:
Table of Contents

Advertisement

DIAGNOSTICS
10DEC05

12941 - Unexpected Retry during Communication to MOTION SYSTEM

DG4825-1
Page

CONTROL BD

84 of 180
Description: A message could not be sent to the SLOW SCAN CONTROLLER
Cause:
Action:

ERASE LAMPS

13001 - 13008 ERASE LAMP PAIRS

Code
13001
13002
13003
13004
13005
13006
13007
13008
Cause:
Action:
BOARD A6. The command is executed again.
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Do the checkout for the
BOARD
A7.
Log Message
ERASE LAMP PAIR No. 1 is out
ERASE LAMP PAIR No. 2 is out
ERASE LAMP PAIR No. 3 is out
ERASE LAMP PAIR No. 4 is out
ERASE LAMP PAIR No. 5 is out
Two ERASE LAMP PAIRS are not functioning
Three ERASE LAMP PAIRS are not functioning
Four ERASE LAMP PAIRS are not functioning
One or more ERASE LAMP PAIRS malfunctioned.
Install a new ERASE LAMP PAIR.
SLOW SCAN CONTROLLER BOARD A6 and COIL
Error Codes

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Directview cr 85056344825

Table of Contents