DIAGNOSTICS
10DEC05
12004 - PLATE PRESENT SENSOR S5 failed to activate
DG4825-1
Page
71 of 180
Cause 1: The PLATE PRESENT SENSOR S5 operates correctly, but another malfunction
Action:
Cause 2: The PLATE PRESENT SENSOR S5 malfunctioned.
Action:
Cause 3: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action:
12009 - Caution! SLOW SCAN MOTOR drawing too much power
Cause:
Action:
occurred.
1.
Check:
•
PLATE was removed from the CASSETTE
•
SLOW SCAN MOTOR operates correctly
2.
Do the checkout for the
BOARD
A7.
3.
Remove any obstructions from the path of the EXTRACTION BAR.
Do the checkout
PLATE PRESENT SENSOR
Do the checkout for the
A malfunction occurred in one or more of the following components:
•
COIL BOARD
•
EXTRACTION BAR SENSOR S9
•
SLOW SCAN AY
1.
Check for an obstruction.
2.
Do the checkout for the
BOARD
A7.
SLOW SCAN CONTROLLER BOARD A6 and COIL
MOTION SYSTEM CONTROL BOARD
SLOW SCAN CONTROLLER BOARD A6 and COIL
S5.
A1.
Error Codes