DIAGNOSTICS
10DEC05
DG4825-1
Page
38 of 180
6
Touch [-100] until the "Set Amplitude" value is 400.
7
Observe the laser beam on the sheet of paper.
8
Touch [+100] until the "Set Amplitude" value is 3000.
9
Observe the laser beam on the sheet of paper.
"Set Amplitude"
10
Does the laser beam change size correctly?
Continue with the next
test.
Laser Beam
400
short
3000
long
Yes
Check the
malfunction.
No
Error Codes
for the cause of the
Using the Diagnostics