Kodak DirectView CR 825 Diagnostics Manual page 65

Hide thumbs Also See for DirectView CR 825:
Table of Contents

Advertisement

DIAGNOSTICS
10DEC05
Cause 2: The internal BAR CODE READER malfunctioned.
DG4825-1
Page
Action:
65 of 180
1.
Check:
alignment of the BAR CODE READER
BAR CODE READER is not blocked
connections to the BAR CODE READER
2.
If the BAR CODE READER emits a beep when you insert a CASSETTE, but
the CASSETTE does not load, check the CONNECTOR RJ45 between the
BAR CODE READER and the MASTER CENTRAL PROCESSING UNIT
BOARD A2.
3.
If the BAR CODE READER does not emit a beep when you load a
CASSETTE, check the CONNECTOR RJ45 between the BAR CODE READER
and the MOTION SYSTEM CONTROL BOARD A1.
Error Codes