32005 - Communications Error Between Mcpu And Ssc Boards - Kodak DirectView CR 825 Diagnostics Manual

Hide thumbs Also See for DirectView CR 825:
Table of Contents

Advertisement

DIAGNOSTICS
10DEC05
Cause 1: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
DG4825-1
Page
Action:
111 of 180
Cause 2: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:

32005 - Communications Error between MCPU and SSC BOARDS

Description: The SLOW SCAN CONTROLLER BOARD A6 did not return a correct
Cause 1: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action:
Cause 2: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:
Do the checkout for the
Do the checkout for the
BOARD
A7.
response to a command from the MOTION SYSTEM CONTROL BOARD
A1. The command is sent again.
Do the checkout for the
Do the checkout for the
BOARD
A7.
MOTION SYSTEM CONTROL BOARD
SLOW SCAN CONTROLLER BOARD A6 and COIL
MOTION SYSTEM CONTROL BOARD
SLOW SCAN CONTROLLER BOARD A6 and COIL
Error Codes
A1.
A1.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Directview cr 85056344825

Table of Contents