DIAGNOSTICS
10DEC05
3
Check that "PASSED" displays for both PMT 1 and PMT 2.
DG4825-1
Page
4
If one of the PMTs displays "FAILED", install a new PMT. Use
40 of 180
AY.
5
If both PMTS display "FAILED", check the connections between:
PMT BOARD
DAS BOARD

GALVO PLOT

Important
The "Galvo Plot" screen has more "Cassette Size" options with Software Version ≥ 4.1.
1
On the "System Service Diagnostic" screen, touch [GALVO PLOT].
2
Touch [START GALVO PLOT].
Using the Diagnostics
PMTs and COLLECTOR