DIAGNOSTICS
10DEC05

15016 - The LASER reference voltage is too high

DG4825-1
Page
93 of 180
Cause:
Action:

15017 - PLATE not calibrated

Cause:
Action:

15018 - There was an error reading or writing to the PCMCIA card

Cause 1: The PCMCIA CARD operates correctly, but another malfunction occurred.
Action:
Cause 2: The PCMCIA CARD malfunctioned.
Action:

15040 - Communications Error Between MCPU and the PC

Description: The TCP/IP SOCKET COMMUNICATIONS were not completed or did
Cause 1: The shutdown of the CR 825/850 SYSTEM was not done in the correct sequence.
Action:
The LASER DIODE in the DETECTOR is backward.
Check:
LASER REFERENCE VOLTAGE in the PMT/DAS BOARD A5
LASER DIODE is installed correctly
The customer scanned a PLATE of a size that was not recognized.
Do:
Calibration for the CR 825/850 SYSTEM
complete the size calibration of the PLATE
Important
A malfunction of the PCMCIA CARD is the last possible cause. Check all other
malfunctions before installing a new PCMCIA CARD.
Check the error logs for the malfunction. Use
Install a new PCMCIA CARD.
not occur between the MASTER CENTRAL PROCESSING UNIT
BOARD A2 and the host computer with Windows 2000.
Use the correct shutdown sequence to de-energize the CR 825/850 SYSTEM.
Checking the Error
Logs.
Error Codes