DIAGNOSTICS
10DEC05
GALVO BOARD A4
DG4825-1
Page
145 of 180
TP1
TP13
TP16
Important
The GALVO BOARD A4:
•
sends signals to control the GALVO MOTOR
•
changes the 24 V DC to:
–
+15 V DC
–
+10 V DC
–
+5 V DC
•
changes –24 to –15 V DC
1
Check for correct voltages on CONNECTOR P2:
+24 V DC
–24 V DC
MULTIMETER
Voltage
Ground
P2-2
P2-2
GALVO BOARD A4
TP15
TP6
TP5
CONNECTOR P2
Positive
P2-1
P2-5
Checkout Procedures
H194_0036BCA
H194_0036BC