Galvo Board A4 - Kodak DirectView CR 825 Diagnostics Manual

Hide thumbs Also See for DirectView CR 825:
Table of Contents

Advertisement

DIAGNOSTICS
10DEC05

GALVO BOARD A4

DG4825-1
Page
145 of 180
TP1
TP13
TP16
Important
The GALVO BOARD A4:
sends signals to control the GALVO MOTOR
changes the 24 V DC to:
+15 V DC
+10 V DC
+5 V DC
changes –24 to –15 V DC
1
Check for correct voltages on CONNECTOR P2:
+24 V DC
–24 V DC
MULTIMETER
Voltage
Ground
P2-2
P2-2
GALVO BOARD A4
TP15
TP6
TP5
CONNECTOR P2
Positive
P2-1
P2-5