Local Panel Error Message Table

This table lists all local panel error message numbers. When an error message is
displayed on the local panel, perform the operator action listed next to the message
number. If the listed action does not clear the error message, call for service.
This table does not include status or action messages. Status messages require no
operator action. Action messages include the required operator action.
Message
Number
P116
Open and close the supply door. Press Supply on the local panel.
P121
Turn the imager OFF, wait five seconds, then turn the imager back
P122
ON.
P123
P126
P138
If pressing Supply does not unlock the door, use the mechanical
release (refer to page 4-14). Darken the room before opening the
door to expose as few sheets of media as possible.
P151
Turn the imager OFF, wait five seconds, then turn the imager back
P154
ON.
P173
Close the platen access door.
P177
Close the cartridge manually (refer to page 4-13).
P178
Close the top cover.
P203
Turn the imager OFF, wait five seconds, then turn the imager back
P204
ON.
P205
P206
P208
Use the mechanical release to open the left door (refer to page
4-14).
P221
Turn the imager OFF, wait five seconds, then turn the imager back
P222
ON.
P223
P224
P225
P226
P227
4-2