DIAGNOSTICS
10DEC05
22011 - MSC BOARD A1: Error during download of MSC Software
DG4825-1
Page
Description: The MOTION SYSTEM CONTROL BOARD A1 received a message
107 of 180
Cause 1: The software did not download.
Action:
Cause 2: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action:
Cause 3: Communications did not occur between:
Action:
MOTION SYSTEM CONTROL BOARD RTXC
29001 - 29008 - MSC BOARD A1: Software malfunctioned
Cause:
Action:
Code
29001 - 29006 System error for the RTXC. The communications data queue is full and
29007
29008
before the preceding message was completed.
1.
Download the software again.
2.
Do the checkout for the
Do the checkout or the
•
MOTION SYSTEM CONTROL BOARD A1
•
MASTER CENTRAL PROCESSING UNIT BOARD A2
Check the RS-232 communications between:
•
MOTION SYSTEM CONTROL BOARD A1
•
MASTER CENTRAL PROCESSING UNIT BOARD A2
The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Do the checkout for the
cannot receive messages. This error should not occur during normal
operation.
System error for the RTXC. This error should not occur during normal
operation.
System error for the MOTION SYSTEM CONTROL. This error should not
occur during normal operation.
MOTION SYSTEM CONTROL BOARD
MASTER CENTRAL PROCESSING UNIT BOARD
MOTION SYSTEM CONTROL BOARD
Description
Error Codes
A1.
A2.
A1.