DIAGNOSTICS
10DEC05
Table 22 Test Points on the SLOW SCAN CONTROLLER BOARD A6
DG4825-1
Page
TP
176 of 180
1
2
14
15
16
17
18
Table 23 SWITCHES on the SLOW SCAN CONTROLLER BOARD A6
SWITCH
S1
Note
The SLOW SCAN CONTROLLER BOARD A6 has no field adjustment points. All adjustments
are preset.
Description
Ground
5 V DC
HALL EFFECT SENSOR
HALL EFFECT SENSOR
HALL EFFECT SENSOR
Ground
12 V DC
Enables and disables the LEDs.
Expected Signal
0 V DC
5.0 V DC
5.0 V DC and 0.0 V DC when the
ROTOR is rotating
5.0 V DC and 0.0 V DC when the
ROTOR is rotating
5.0 V DC and 0.0 V DC when the
ROTOR is rotating
0 V DC
12.0 V DC
Description
Test Points and LEDs