DIAGNOSTICS
10DEC05
Cause:
DG4825-1
Page
Action:
106 of 180
22009 - MSC BOARD A1: Error during download of MSC Software
Description: The boot program for the MOTION SYSTEM CONTROL BOARD A1
Cause:
Action:
22010 - MSC BOARD A1: Error during download of MSC Software
Description: The boot program for the MOTION SYSTEM CONTROL BOARD A1
Cause 1: The software did not download.
Action:
Cause 2: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action:
Cause 3: Communications did not occur between:
Action:
The software did not download.
1.
Download the software again.
2.
Do the checkout for the
3.
If this error continues, install a new MOTION SYSTEM CONTROL BOARD A1.
received a signal that is not correct.
The software did not download.
1.
Download the software again.
2.
Do the checkout for the
3.
If this error continues, install a new MOTION SYSTEM CONTROL BOARD A1.
received a command that was not correct.
1.
Download the software again.
2.
Do the checkout for the
Do the checkout for the
MOTION SYSTEM CONTROL BOARD A1
MASTER CENTRAL PROCESSING UNIT BOARD A2
Check the RS-232 communications between:
MOTION SYSTEM CONTROL BOARD A1
MASTER CENTRAL PROCESSING UNIT BOARD A2
MOTION SYSTEM CONTROL BOARD
MOTION SYSTEM CONTROL BOARD
MOTION SYSTEM CONTROL BOARD
MASTER CENTRAL PROCESSING UNIT BOARD
Error Codes
A1.
A1.
A1.
A2.

Hide quick links: