DIAGNOSTICS
10DEC05
10022 - MSC BOARD A1: Software malfunction
DG4825-1
Page
Description: The DRIVE ROLLER received a signal when waiting for an interrupt or
64 of 180
Cause 1: The software for the MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action:
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action:
10024 - MSC BOARD A1: Software malfunction
Description: A timeout occurred setting up the motion for the DRIVE ROLLER.
Cause 1: The software for the MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action:
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action:
10029 - READER detected the CASSETTE BAR CODE label as backward
Cause 1: The CASSETTE was inserted backward into the CR 825/850 SYSTEM.
Action:
Cause 2: The BAR CODE LABEL is installed backward on the CASSETTE.
Action:
10030 - READER did not read the CASSETTE BAR CODE
Cause 1: The CASSETTE:
Action:
timeout to occur.
Install the software again for the MOTION SYSTEM CONTROL BOARD A1.
Do the checkout for the
Install the software again for the MOTION SYSTEM CONTROL BOARD A1.
Do the checkout for the
Insert the CASSETTE correctly.
1.
Remove the BAR CODE LABEL.
2.
Install a new BAR CODE LABEL.
•
has a bar code that you cannot read
•
does not have a bar code
Install the correct bar code.
MOTION SYSTEM CONTROL BOARD
MOTION SYSTEM CONTROL BOARD
Error Codes
A1.
A1.