10DEC05

19005 - Check the GALVO reference (too low)

DG4825-1
Page
Cause:
102 of 180
Action:

19006 - Memory self test failed. Replace the MCPU BOARD

Cause:
Action:

19009 - The LASER diode off power is too high

Cause 1: The GALVO BOARD A4 malfunctioned.
Action:
Cause 2: The LASER DRIVER BOARD A17 malfunctioned.
Action:
Cause 3: The CABLE between the LASER DRIVER BOARD A17 and the GALVO BOARD
Action:

19010 - The LASER diode on power is too low

Cause 1: The GALVO BOARD A4 malfunctioned.
Action:
Cause 2: The LASER DRIVER BOARD A17 malfunctioned.
Action:
Cause 3: The CABLE between the LASER DRIVER BOARD A17 and the GALVO BOARD
Action:
The reference voltage for the GALVO AGC is too low.
Check that the R2 AGC signal is correctly adjusted.
The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Install a new MASTER CENTRAL PROCESSING UNIT BOARD A2.
Do the checkout for the
Do the checkout for the
A4 malfunctioned.
Check the CABLE.
Do the checkout for the
Do the checkout for the
A4 malfunctioned.
Check the CABLE.
GALVO BOARD
A4.
LASER DRIVER BOARD
GALVO BOARD
A4.
LASER DRIVER BOARD
A17.
A17.