DIAGNOSTICS
10DEC05
Cause 2: The CASSETTE LOAD SENSOR S2 malfunctioned.
DG4825-1
Page
Action:
63 of 180
10020 - Malfunction at CASSETTE ENTRY SENSOR S1
Description: The CASSETTE ENTRY SENSOR S1 is blocked when it should be clear.
Cause 1: The CASSETTE ENTRY SENSOR S1 malfunctioned.
Action:
10021 - MSC BOARD A1: Software malfunction
Description: The DRIVE ROLLER received a command that was not correct.
Cause 1: The software for the MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action:
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action:
Check:
•
status of the CASSETTE LOAD SENSOR S2 - use
Control
•
LED DS21 on the MOTION SYSTEM CONTROL BOARD A1 changes status
•
power to the CASSETTE LOAD SENSOR S2
•
condition of the EMITTER and DETECTOR of the CASSETTE LOAD SENSOR
S2
Check:
•
status of the CASSETTE ENTRY SENSOR S1 - use
Control
•
LED DS19 on the MOTION SYSTEM CONTROL BOARD A1 changes status
•
power to the CASSETTE ENTRY SENSOR S1
•
condition of the EMITTER and DETECTOR of the CASSETTE ENTRY
SENSOR S1
Install the software again for the MOTION SYSTEM CONTROL BOARD A1.
Do the checkout for the
MOTION SYSTEM CONTROL BOARD
Error Codes
Individual Component
Individual Component
A1.