Kodak DirectView CR 825 Diagnostics Manual page 150

Hide thumbs Also See for DirectView CR 825:
Table of Contents

Advertisement

DIAGNOSTICS
10DEC05
3
Check the voltage on the SLOW SCAN CONTROLLER BOARD A6:
DG4825-1
Page
150 of 180
4
If the voltage is not correct, do the checkout
5
Check the voltage on the SLOW SCAN CONTROLLER BOARD A6:
6
On the "System Service Diagnostics" screen, touch [Display Software Versions].
7
Check for the correct software versions for the MOTION SYSTEM CONTROL BOARD A1
and the SLOW SCAN CONTROLLER BOARD A6.
8
If the screen displays "Not Responding", check that the RS-232 communications on the
CABLE A6J5 operate correctly.
9
Download the software to the SLOW SCAN CONTROLLER BOARD A6.
10
If you cannot successfully download the software, install a new SLOW SCAN
CONTROLLER BOARD A6.
11
During the time the equipment is energized, disconnect the CONNECTOR J1 from the
SLOW SCAN CONTROLLER BOARD A6.
Note
Disconnecting the CONNECTOR J1 allows you to manually rotate the LEAD SCREW on the
SLOW SCAN AY.
MULTIMETER
Negative
Positive
TP1
TP18
MULTIMETER
Negative
Positive
TP1
TP2
Voltage
+24 V DC
INTERLOCK SWITCH
Voltage
+5 V DC
Checkout Procedures
S8.