Page

15001 - MCPU BOARD A2: Fatal Error in Software

89 of 180
Description: A PSOS system error occurred that it is not possible to repair. The
Cause:
Action:
15002 - MCPU BOARD A2: Fatal Error in Software
Description: A system software error occurred. The "Event Log" will include the PSOS
Cause:
Action:

15003 - MCPU BOARD A2: Communications Error in Software

Description: A timeout occurred for a communication between the MASTER
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action:
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action:

15005 - LASER: Reference Error

Description: The reference voltage of the LASER is less than 50% of the original
"Event Log" includes the PSOS system error number, module, and line
number.
The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Do the checkout for the
system error number, module, and line number.
The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Do the checkout for the
CENTRAL PROCESSING UNIT BOARD A2 and the MOTION SYSTEM
CONTROL BOARD A1.
Do the checkout for the
Do the checkout for the
value.
MASTER CENTRAL PROCESSING UNIT BOARD
MASTER CENTRAL PROCESSING UNIT BOARD
MASTER CENTRAL PROCESSING UNIT BOARD
MOTION SYSTEM CONTROL BOARD
Error Codes
A2.
A2.
A2.
A1.