Error Messages

6–4
TROUBLESHOOTING
Error Messages
The following table lists the error messages and the actions to take.
Table 2
Error Messages
Error
Message
The radiography control button was released
E01 + audible
before the exposure end. The display indicates
alarm
the remaining exposure time. Based on this time,
you must decide whether to develop the film or
make another exposure).
The radiography control was enabled while the
E02
unit was being powered ON.
E03-E04
Exposure time control problems
E10 to E18
kV voltage error
E20 to E24
Filament voltage error
Problem with voltage to the main power supply
E30
or to the chemical capacitor
System error (problems with the power board
E40 to E46
microprocessor)
Problem with the IC bus, the connection
E50 to E54
between the control panel and the power board.
Possible Cause
Action
Quickly press on the selection knob
to stop the alarm.
Switch OFF the unit, then restart it. If
the problem persists, contact your
representative and stop using the
equipment.