16006 - Calibration Error: Did Not Acquire Image; 16010 - Possible Light Leak; 16011 - Imaging System Error Occurred During Initialization; 16012 - Pmt/Das Board A5: Calculation Error Occurred During - Kodak DirectView CR 825 Diagnostics Manual

Hide thumbs Also See for DirectView CR 825:
Table of Contents

Advertisement

DIAGNOSTICS
10DEC05

16006 - Calibration Error: Did not acquire image

DG4825-1
Page
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
97 of 180
Action:
Cause 2: The DIGITIZER BOARD A3 malfunctioned.
Action:
Cause 3: The GALVO BOARD A4 malfunctioned.
Action:

16010 - Possible Light Leak

Description: Large quantities of light occurred during the "AUTO-ZERO."
Cause:
Action:

16011 - Imaging System Error occurred during Initialization

Cause 1: The PMT/DAS BOARD A5 malfunctioned.
Action:

16012 - PMT/DAS BOARD A5: Calculation Error occurred during

Initialization

Description: The value for one or more of the following parameters did not match the
Do the checkout for the
Do the checkout for the
Do the checkout for the
The DOOR is open, a PANEL is loose, or a light leakage occurred in the
CASSETTE HANDLING SYSTEM.
1.
Do the
Light Test
2.
Repair the light leakage.
1.
Check the CABLE between the PMT BOARD and the DAS BOARD.
2.
Do:
checkout for the
MCPU Self Test
specification when the CR 825/850 SYSTEM initialized: "Fine or Coarse
Offset", or "Fine or Coarse Ratio."
MASTER CENTRAL PROCESSING UNIT BOARD
DIGITIZER BOARD
GALVO BOARD
A4.
to locate the source of the light leakage.
PMT/DAS BOARD A5
A3.
Error Codes
A2.