DIAGNOSTICS
10DEC05
16006 - Calibration Error: Did not acquire image
DG4825-1
Page
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
97 of 180
Action:
Cause 2: The DIGITIZER BOARD A3 malfunctioned.
Action:
Cause 3: The GALVO BOARD A4 malfunctioned.
Action:
16010 - Possible Light Leak
Description: Large quantities of light occurred during the "AUTO-ZERO."
Cause:
Action:
16011 - Imaging System Error occurred during Initialization
Cause 1: The PMT/DAS BOARD A5 malfunctioned.
Action:
16012 - PMT/DAS BOARD A5: Calculation Error occurred during
Initialization
Description: The value for one or more of the following parameters did not match the
Do the checkout for the
Do the checkout for the
Do the checkout for the
The DOOR is open, a PANEL is loose, or a light leakage occurred in the
CASSETTE HANDLING SYSTEM.
1.
Do the
Light Test
2.
Repair the light leakage.
1.
Check the CABLE between the PMT BOARD and the DAS BOARD.
2.
Do:
•
checkout for the
•
MCPU Self Test
specification when the CR 825/850 SYSTEM initialized: "Fine or Coarse
Offset", or "Fine or Coarse Ratio."
MASTER CENTRAL PROCESSING UNIT BOARD
DIGITIZER BOARD
GALVO BOARD
A4.
to locate the source of the light leakage.
PMT/DAS BOARD A5
A3.
Error Codes
A2.