Cam Sensors S10 And S11 - Kodak DirectView CR 825 Diagnostics Manual

Hide thumbs Also See for DirectView CR 825:
Table of Contents

Advertisement

DIAGNOSTICS
10DEC05

CAM SENSORS S10 and S11

DG4825-1
Page
137 of 180
1
Check the condition of:
SENSOR S10
SENSOR S11
2
Measure the voltage at SENSOR S10:
3
If either measurement is not correct, check the HARNESS P2/J2.
4
Measure the voltage at SENSOR S11:
5
If either measurement is not correct, check the HARNESS P2/J2.
SENSOR S11
DUPLEX CAM
H194_1033ACA
H194_1033AC
MULTIMETER
Negative
Positive
S10-5
S10-3
MULTIMETER
Negative
Positive
S11-5
S11-1
DUPLEX CAM ENCODER
BOARD A8
SENSOR S10
Status of
Measurement
SENSOR S10
Blocked
Not Blocked
< 0.5 V DC
Status of
Measurement
SENSOR S11
Blocked
Not Blocked
< 0.5 V DC
Checkout Procedures
LED DS26
5 V DC
Illuminated
Not illuminated
LED DS27
5 V DC
Illuminated
Not illuminated