DIAGNOSTICS
10DEC05

55110 - Could not determine what product we are running on

DG4825-1
Page
118 of 180
Cause:
Action:

55111 - The Options Diskette is not valid for this product

Cause:
Action:

55112 - Error obtaining Adapter Name on Pentium 4 class machine

Cause:
Action:

55113 - Error obtaining MAC address for card name specified in

AdapterName in General Equipment on Pentium 4 class machine

Cause:
Action:
The correct configuration for this CR 825/850 SYSTEM was not provided.
1.
Boot the SYSTEM.
2.
Install the OPTIONS DISKETTE again.
The OPTIONS DISKETTE is not correct for the CR 825/850 SYSTEM.
1.
Obtain the correct OPTIONS DISKETTE.
2.
Install the OPTIONS DISKETTE again.
Note
This error does not display on a Pentium III PROCESSOR.
The correct configuration for this CR 825/850 SYSTEM was not provided.
1.
Boot the SYSTEM.
2.
Install the OPTIONS DISKETTE again.
Note
This error does not display on a Pentium III PROCESSOR.
The correct configuration for this CR 825/850 SYSTEM was not provided.
1.
Boot the SYSTEM.
2.
Install the OPTIONS DISKETTE again.
Error Codes