Self Test" For The Master Central Processing Unit; 19001 - Check Galvo Communications Cable; 19002 - Check Pmt Board Cable And Power; 19003 - Check The Power To The Galvo Board - Kodak DirectView CR 825 Diagnostics Manual

Hide thumbs Also See for DirectView CR 825:
Table of Contents

Advertisement

DIAGNOSTICS
10DEC05
Cause 4: The PMT/DAS BOARD A5 malfunctioned.
DG4825-1
Page
Action:
101 of 180

"Self Test" for the MASTER CENTRAL PROCESSING UNIT

19001 - Check GALVO COMMUNICATIONS CABLE

Cause:
Action:

19002 - Check PMT BOARD CABLE and power

Cause:
Action:

19003 - Check the power to the GALVO BOARD

Cause:
Action:
19004 - GALVO AGC reference is too high. Suggest check if GALVO

connected and/or adjust AGC

Cause:
Action:
Do the checkout for the
Communication problems in the GALVO AY.
Check the GALVO COMMUNICATION CABLE.
The PMT/DAS BOARD A5 malfunctioned.
Check:
CABLE to the PMT/DAS BOARD A5
power to the PMT/DAS BOARD A5
PMT/DAS BOARD A5
No power to the GALVO BOARD A4.
Check the power to the GALVO BOARD A4.
The reference voltage for the GALVO AGC is too high.
Check:
connection to the GALVO BOARD A4
signal for the R2 AGC is correctly adjusted
connections at the GALVO AY
PMT/DAS BOARD
A5.