DIAGNOSTICS
10DEC05

15007 - Communications Error between MCPU BD A2 and GALVO BD A4

DG4825-1
Page
Description: The DMA process for the MASTER CENTRAL PROCESSING UNIT
91 of 180
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action:
Cause 2: The DIGITIZER BOARD A3 malfunctioned.
Action:
Cause 3: The GALVO BOARD A4 malfunctioned.
Action:

15008 - Start of Scan Timeout

Description: A timeout occurred for the expected "START OF SCAN".
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action:
Cause 2: The DIGITIZER BOARD A3 malfunctioned.
Action:
Cause 3: The GALVO BOARD A4 malfunctioned.
Action:

15012 - Communications Error between MCPU BOARD A2 and the PC

Description: Windows 2000 did not return a response to a message within a given
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action:
Cause 2: The TCP/IP connection malfunctioned.
Action:
BOARD A2 did not complete within the expected time.
Do the checkout for the
Do the checkout for the
Do the checkout for the
Do the checkout for the
Do the checkout for the
Do the checkout for the
time.
Do the checkout for the
Check the TCP/IP connection to the computer. See
MASTER CENTRAL PROCESSING UNIT BOARD
DIGITIZER BOARD
A3.
GALVO BOARD
A4.
MASTER CENTRAL PROCESSING UNIT BOARD
DIGITIZER BOARD
A3.
GALVO BOARD
A4.
MASTER CENTRAL PROCESSING UNIT BOARD
Step
6,
Page
10.
Error Codes
A2.
A2.
A2.