DIAGNOSTICS
10DEC05
12002 - EXTRACTION BAR REFERENCE SENSOR S9 failed to activate
DG4825-1
Page
70 of 180
Cause 1: The EXTRACTION BAR SENSOR S9 operates correctly, but another malfunction
Action:
Cause 2: The EXTRACTION BAR SENSOR S9 malfunctioned.
Action:
12003 - PLATE PRESENT SENSOR S5 did not detect PLATE
Cause 1: The PLATE PRESENT SENSOR S5 operates correctly, but another malfunction
Action:
Cause 2: The PLATE PRESENT SENSOR S5 malfunctioned.
Action:
occurred.
1.
Remove any obstructions from the path of the EXTRACTION BAR.
2.
Do the checkout for the
BOARD
A7.
Do the checkout for the
occurred.
1.
Check:
•
PLATE was removed from the CASSETTE
•
SLOW SCAN MOTOR operates correctly
2.
Do the checkout for the
BOARD
A7.
3.
Remove any obstructions from the path of the EXTRACTION BAR.
Do the checkout
PLATE PRESENT SENSOR
SLOW SCAN CONTROLLER BOARD A6 and COIL
EXTRACTION BAR SENSOR
SLOW SCAN CONTROLLER BOARD A6 and COIL
S9.
S5.
Error Codes