DIAGNOSTICS
10DEC05
Calibration
DG4825-1
Page
16001 - Calibration Error: Did not find START OF SCAN position
95 of 180
Description: During the calibration, the CR 825/850 SYSTEM did not locate the
Cause 1: Calibration of the CR 825/850 SYSTEM is not correct.
Action:
Cause 2: The alignment of the optics is not correct.
Action:
Cause 3: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:
16002 - Calibration Error: Did not find Offset
Description: The GALVO did not detect the "Offset".
Cause 1: The exposure of the PLATE is not correct.
Action:
Cause 2: The GALVO is not aligned correctly.
Action:
16003 - Calibration Error: Did not find Amplitude
Description: The GALVO BOARD A4 did not detect the "Amplitude".
"START OF SCAN" position within the given time.
Do the
Calibration for the CR 825/850
Do:
•
adjustment for the
•
adjustment of the
Do the checkout for the
BOARD
A7.
1.
Use the DOSIMETER to check the exposure of the PLATE.
2.
Check for scratches on the edges of the SCREEN.
Do:
•
adjustment of the
•
checkout for the
GALVO BOARD A4
SYSTEM.
FOLD MIRROR
GALVO ROTATION
SLOW SCAN CONTROLLER BOARD A6 and COIL
GALVO ROTATION
Error Codes