Kodak DirectView CR 825 Diagnostics Manual page 34

Hide thumbs Also See for DirectView CR 825:
Table of Contents

Advertisement

DIAGNOSTICS
10DEC05
4
Wait for the test to complete.
DG4825-1
Page
34 of 180
5
Check that the value is:
5 -180 between "Start of Scan" and "End of Scan"
< 660 between "End of Scan" and "End of Travel"
no spikes > 30
6
If the results are not correct, check the error logs. See
7
Touch [START VELOCITY TEST].
8
Wait for the test to complete.
9
Check that the results between "Start of Scan" and "End of Scan" are below 90 and
stable.
Using the Diagnostics
Checking the Error
Logs.