10009 - Cassette Load Sensor S2 Did Not Clear - Kodak DirectView CR 825 Diagnostics Manual

Hide thumbs Also See for DirectView CR 825:
Table of Contents

Advertisement

DIAGNOSTICS
10DEC05

10009 - CASSETTE LOAD SENSOR S2 did not clear

DG4825-1
Page
Cause 1: The CASSETTE LOAD SENSOR S2 operates correctly, but an obstruction
60 of 180
Action:
Cause 2: The CASSETTE LOAD SENSOR S2 malfunctioned.
Action:
occurred.
Check:
DRIVE ROLLERS rotate freely
CASSETTE LOAD SENSOR S2 for obstructions
adjustment of the
Check:
status of the CASSETTE LOAD SENSOR S2 - use
Control
LED DS21 on the MOTION SYSTEM CONTROL BOARD A1 changes status
power to the CASSETTE LOAD SENSOR S2
condition of the EMITTER and DETECTOR of the CASSETTE LOAD SENSOR
S2
SLED CAM FOLLOWER