DIAGNOSTICS
10DEC05
Cause 1: The exposure of the PLATE is not correct.
DG4825-1
Page
Action:
96 of 180
Cause 2: The GALVO is not aligned correctly.
Action:
16005 - Calibration Error: Routine did not receive image in the time
allowed
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action:
Cause 2: The DIGITIZER BOARD A3 malfunctioned.
Action:
Cause 3: The GALVO BOARD A4 malfunctioned.
Action:
1.
Use the DOSIMETER to check the exposure of the PLATE.
2.
Check for scratches on the edges of the SCREEN.
Do:
•
adjustment of the
•
checkout for the
GALVO BOARD A4
Do the checkout for the
Do the checkout for the
Do the checkout for the
GALVO ROTATION
MASTER CENTRAL PROCESSING UNIT BOARD
DIGITIZER BOARD
GALVO BOARD
A4.
A3.
Error Codes
A2.