Kodak DirectView CR 825 Diagnostics Manual page 13

Hide thumbs Also See for DirectView CR 825:
Table of Contents

Advertisement

DIAGNOSTICS
10DEC05
Table 2 Calibration
DG4825-1
Page
13 of 180
[View Collector Profiles]
[Scan Calibration]
BUTTON
For the procedure, see
Displays the COLLECTOR PROFILE for each size of CASSETTE,
including:
number of steps from "HOME" to "START OF SCAN"
"OFFSET" - Number of pixels from "START" to the diagnostic
image
"Amplitude" - Number of pixels from "OFFSET" to the last pixel
of the diagnostic image
PMT 1 - 2: Correction value of each PMT for gain adjustment
All COVERS must be installed when using this option.
For the procedure, see
SYSTEM.
Completes the calibration of:
Starts a COLLECTOR PROFILE for all sizes of STORAGE
PHOSPHOR PLATES.
Allows calibration of the PLATES exposed at 20 mR ± 5 mR.
Displays the date and time of the last successful calibration.
Important
Calibration for the CR 825/850
Caution
GALVO
PMT 1
PMT 2
Description
Calibration for the CR 825/850
Using the Diagnostics
SYSTEM.